Jitter measurement by Intramıuscular Axonal Microstimulation
Levent Sinan BİR, Reha KURUOĞLU
Gazi Üniversitesi Tıp Fakültesi Nöroloji Anabilim Dalı, Ankara
Keywords: Axonal microstimulation, single fiber electromyography, extensor digitorum communis, jitter.
Abstract
Single fiber electromyography (SFEMG) is a sensitive diagnostic tool for neuromusculartransmission disorders. Stimulation SFEMG is generally preferred, due to the ease qf theprocedure. By using this method, we investigated the extensor digitorum communis muscle in 10 healthy volunteers. Compared with the study in voluntarily activated muscle, we found a significantly lower jitter. Our results suggest that the upper normal limit for the mean jitter is 23 µsec, while it is 38 µsec far the jitter of the individual motor end plates.